Written By Renata Roberts –
Pharmaceutical manufacturers of packaging system for pharmaceutical drugs must ensure that their packaging system does not react physically or chemically with the drug product so to alter its safety, identity, strength, quality, or purity according to 21 CFR 211.94(a). Proper characterization of the materials of construction for the packaging system must be performed to ensure its suitability for its intended use.
Within the USP, chapter <661.1>, Plastic Materials of Construction, provides tests and rationale for the chemical characterization of plastic materials of construction used for packaging components and systems.
Within USP <661.1>, verification of material Identity is a critical aspect of characterization, and two analytical techniques are proposed for its evaluation: Fourier-Transform Infrared Spectroscopy (FTIR) and Differential Scanning Calorimetry (DSC).
The chapter includes many reference standards for common plastic types, such as HDPE, LDPE, PP homopolymer, and plasticized PVC, but is not all inclusive. In some cases, a reference material or reference scan is required to confirm identity of the material to manufacturer's intent.
In some cases, both Identification by IR and by Thermal Analysis (DSC) are permitted for a specific plastic type, however, only one is required to verify the identity of the material. In this blog, we will explore what may impact the choice to choose one identification technique over the other.
What is FTIR Methodology
Generally, FTIR works on the principle of light absorption over a wavelength range. As each unique material will have a different arrangement and concentration of molecular bonds and functional groups, each material absorbs infrared light differently. Thus, different materials have a unique spectrum that functions as a spectral fingerprint, which can be compared to known reference standards to confirm identify of a given material.
In USP <661.1>, FTIR is used in the identification test for the following materials of construction:
- Cyclic Olefins
- Polyamide 6
- Polycarbonate
- High density polyethylene (HDPE)
- Low density polyethylene (LDPE)
- Polyethylene terephthalate (PET) and polyethylene terephthalate G (PETG)
- Poly(Ethylene-Vinyl Acetate) (P-EVA)
- Polypropylene (PP) – homopolymer only
- Polyvinyl Chloride (PVC)
- Polyvinyl Chloride Plasticized (PVC, plasticized)
What is the DSC Methodology
A differential scanning calorimeter (DSC) is an instrument used to measure and characterize melting points and heat flow of different materials. Similar to FTIR, technique can be used across material types and is not specific to plastic packaging. USP <891> describes the technique generally, and how it can be used in a sample-agnostic way. It is referenced in USP <661.1>, where it is applied to plastic materials of construction.
By assessing a known mass of material to a thermal analysis cycle, and comparing melting point and thermal curve to that of a reference standard, the identity of the material can be determined.
The DSC analysis tells us: thermal behavior of materials. This information is relevant because thermal behavior of molecules can be used to differentiate the broader polymer family because of differences in molecular weight, crystallinity, additives, processing conditions, cooling history and manufacturing conditions. USP <661.1>, DSC is used in the identification test for the following materials of construction:
- Polyamide 6
- Polycarbonate
- Polyethylene
- Polyethylene terephthalate and polyethylene terephthalate G
- Poly(Ethylene-Vinyl Acetate)
- Polypropylene
- Polyvinyl Chloride
- Polyvinyl Chloride Plasticized
FTIR VS DSC
FTIR uses the absorption of light across wavelengths to generate a unique material-specific fingerprint, while DSC measures the thermal behavior of a material as it is heated and cooled.
Under USP <661.1>, DSC is not a method used for the characterization of cyclic olefins. The explanation is given in the USP <661.1> Revision Bulletin which states given the amorphous nature of the polymers and copolymers, material-to-material variations in the transition temperature (Tg ) can be anticipated; therefore, it is neither recommended nor required that differential scanning calorimetry be performed for this material.
However, in reality, this type of scenario can occur with other plastic types, including blended materials, such as MDPE, or PP copolymers, as well as materials that are unaddressed by the chapter, where no heating/cooling curve (method) is proposed by the chapter.
Furthermore, where material suppliers almost always have FTIR reference scans for their materials, there is no such common practice for thermal analysis results. This enables comparison to reference scans under FTIR that would not be possible under DSC for those materials not addressed by the chapter, and for which no USP RS exists. Given the wide array of materials actually used in packaging, relative to those covered by the standard alone, FTIR often winds up being the most effective, efficient, and industry-standard approach to verifying material identity for plastic packaging systems.